Specifications
Materials of construction
Filter Media: Supor™ membrane [hydrophilic polyethersulfone (PES)]
Serum Acrodisc™ Prefilter Media: Binder-free borosilicate glass
Acrodisc Housing:
13 mm: Polypropylene
25 and 32 mm: Modified acrylic
37 mm: Polypropylene
Serum Acrodisc Housing: ABS
Pore Size
0.1, 0.2, 0.45, 0.8, 1.2, and 5 µm
Effective Filtration Area
13 mm: 1.0 cm 2
25 mm: 2.8 cm 2
32 mm: 5.8 cm 2
37 mm: 7.5 cm 2
Inlet/Outlet Connections
Female luer lock inlet, male slip luer outlet
Typical Hold-Up Volume
(with air purge)
13 mm: < 28 µL
25 mm: < 70 µL
32 and 37 mm: <100 µL
Maximum Operating Temperature
55°C (131°F)
Maximum Operating Pressure
5.2 bar (520 kPa, 75 psi)
Typical Water Flow Rate
mL/min at 3.1 bar (310 kPa, 45 psi)
0.1 μm, 25 mm: 35
0.1 μm, 32 mm: 100
0.2 μm, 13 mm: 22
0.2 μm, 25 mm: 175
0.2 μm: 22 0.2 μm, 25 mm: 175
0.2 μm , 30. , 13 mm: 35
0.45 μm, 25 mm: 300
0.45 μm, 32 mm: 700
0.8 μm, 13 mm: 150
0.8 μm, 25 mm: 700
0.8
/0.2 μm, 25 mm: 145
0.8/3. mm: 1700
5 μm, 32 mm: 1750
GF/0.2 μm, 37 mm (Serum Acrodisc Filter): 425
Endotoxin Level
13mm, 25mm and 37mm: < 0.25 EU/mL using Limulus Amoebocyte Lysate (LAL)
32mm: <0.5EU/mL using Limulus Amoebocyte Lysate (LAL) test
Sterilization
Sterilized by gamma irradiation and individually blister packaged.
Applications
0.1 and 0.2 μm pore sizes provide sterilization of small volumes of buffers, culture media, and additives.
Acrodisc PF and Serum Acrodisc syringe filters are ideal for clarification/sterilization ofalidal seringe filters are ideal for clarification/sterilization ofalidal orityilization mot
. for prefiltration and particulate removal.