Acrodisc™ syringe filters with PTFE membrane, a Pall Life Sciences product, provide exceptional chemical and temperature compatibility.
Ideal for filtration of gas and/or organic solvents.
Prevents spurious peaks on chromatograms for accurate experimental results. HPLC certified for low levels of UV-absorbing extractables.
Prevents “air bubble” lock when filling sample vials with 13 mm minispike outlet.
4 mm device offers the lowest sample hold-up (< 10 μL) and is useful for samples up to 2 mL in volume.
Specifications
Materials of Construction
Filter Media: Hydrophobic PTFE membrane on a polypropylene support
Housing: Polypropylene
GxF Prefilter: Borosilicate glass
4, 13, and 25 mm PSF: Female luer lock inlet, male slip luer outlet
13 mm: Available with minispike outlet
Typical Hold-Up Volume (with air purge)
4 mm: < 10 μL
13 mm (minispike): < 14 μL
13 mm (male luer): < 30 μL
25 mm PSF: < 125 μL
Typical Liquid Flow Rate (MeOH)
mL/min at 1.0 bar (100 kPa, 15 psi)
25 mm PSF, 0.2 μm: 245
25 mm PSF, 0.45 μm: 510
25 mm PSF, GxF/0.45 μm: 395
Maximum Operating Temperature
100°C (212°F) at 2.1 bar (210 kPa, 30 psi)
Maximum Operating Pressure
4 mm: 5.2 bar (520 kPa, 75 psi)
13 mm: 6.9 bar (690 kPa,100 psi)
25 mm PSF: 4.1 bar (410 kPa, 60 psi) at 21 – 24°C (70 – 75°F); 2.1 bar (210 kPa, 30 psi) at 100°C (212°F)
Sterilization of 25 mm Filters
Provided non-sterile. If desired, autoclave at 121- 123°C (250 – 253°F) at 1.0 bar (100 kPa, 15 psi) for a maximum of 15 min.
Applications
Ultimate in chemical compatibility for filtering harsh chemicals that destroy other membrane materials.
The Acrodisc PSF GxF syringe filter provides two to four times the throughput of standard prefilter devices.
Moisture barrier for venting applications.
25 mm Acrodisc PSF syringe filter is available in robotic-compatible AutoPack™ packaging